PhosphoSens Product Support
How Can We Help?
PhosphoSens-Kinetic Kinase Assay
Reaction setup and protocols
PhosphoSens-Red Kinase Assay
Reaction setup and protocols
PhosphoSens-Kinetic Phosphatase Assay
Reaction setup and protocols
Assay Validation Reports
Detailed presentations of assay validation experiments
Frequently Asked Questions
General and troubleshooting
Contact Our Team
Reach out to our team for additional information or support
PhosphoSens-Kinetic General Plate Reader Setup
- Use an instrument capable of reading fluorescence intensity in kinetic (continuous) mode at an excitation wavelength of ~360 nm (358-363 nm) and an emission wavelength of ~492 nm (485-498 nm).
- Readings can be made at your desired intervals and duration. We read every 30 seconds for 4 hours (or until assay plateaus).
- All PhosphoSens Assays are developed and optimized to be run at 30°C.
- Place your plate in the reader, and select a well to use for adjusting gain and focus. To adjust the gain, use a positive control (phosphopeptide) to avoid going off scale during the assay. When finished, select "Start Adjustment" (or equivalent) to calculate its optimal focal height and make any gain adjustments necessary.
PhosphoSens-Red General Plate Reader Setup
- Use an instrument capable of reading an endpoint time-resolved fluorescence (RFU) reading with an excitation wavelength of ~360 nm and an emission wavelength of ~620 nm.
- After the kinase reaction is complete, and Europium has been added, return your plate to the reader without it's plate seal, adjust the gain and read.
Plate Reader-Specific Setup Guides
For additional instrument-specific setup support, contact our support team.
BMG LABTECH PHERAstar/PHERAstarPlus /PHERAstar FS/PHERAstar FSX Microplate Readers Assay Setup Guide
General Questions
How many assays can I run per 1 mg of Sensor Peptide?
PhosphoSens Sensor Peptides range in molecular weight (MW) based on the size and sequence of the substrate. The average MW or a sensor peptide is 2,150 g/mol. This size Sensor Peptide will be re-suspended in approximately 450 µL of diluent. If running an assay in a 384-well plate, with a final well volume of 20 µL and Sensor Peptide concentration of 10 µM, you can run 2,250 assays!
How do I resuspend my lyophilized sensor peptide?
Reference the "Solubilization" section of the printed Certificate of Analysis (CofA) that is printed and shipped with every order. If you do not see this, please contact us at support@assayquant.com and we will share an electronic copy of your CofA.
How should I store my sensor peptide?
Lyophilized sensor peptides can be stored up to 2 years at -20°C or below.
Once resuspended, sensor peptides should be aliquoted and stored at -20°C for up to 1 year. Minimize thaw/freeze cycles for best stability.
How do I place an order?
Orders can be placed using the following methods:
1. If you know what you'd like to order, contact your Area Business Manager or orders@assayquant.com.
2. If you are a new customer or have any questions, please contact us at hello@assayquant.com.
Additional Resources
Looking for more? Check out our Resources for all the latest information.
Kinome Tree Mapping Application
Visualize, explore, and customize your kinome profiling results with our NEW Kinome Tree Mapping Application!
Assay Set Up Questions
What plates should I use?
At AssayQuant, we have validated the following plates:
- Corning 96-well Half Area White Flat Bottom Polystyrene NBS Microplate (Product Number #3642)
- Corning Low Volume 384-well White Flat Bottom Polystyrene NBS Microplate (Product Number #3824)
- ProxiPlate 384-shallow well Plus, White (Product Number #6008280)
Are all assays performed under the same reaction conditions?
While most of our assays use standard reaction conditions (54 mM HEPES, pH 7.5, 1 mM or KmATP, 1.2 mM DTT, 0.55 mM EGTA, 0.012% Brij-35, 10 mM MgCl2, 10 µM sensor peptide substrate, 0.05 - 10 nM kinase) we carefully optimize each assay during development to ensure that co-factors or additives are considered. For more information, please visit our Product Support Pages and look at the “Target-Specific Reaction Conditions Dropdown” or reach out to us at Support@assayqunat.com for additional information.
Does enzyme source matter?
Yes! We see significant differences in levels of activity between enzyme vendors. When choosing a commercially available kinase preparation, an assessment of purity, specific activity, the nature and location of co-expression and purification tags (e.g., N- or C-terminal GST, His, or FLAG tags), and the size of the construct (full-length or truncated) should be considered. The most rigorous approach is to obtain a kinase from multiple sources and compare the activity with the PhosphoSens platform, where the kinetic format allows a quantitative measurement allowing the most appropriate enzyme to be selected for further study.
Can you exclude both BSA and Brij-35 in the reaction conditions?
BSA and Brij-35 both work to stabilize the enzyme and prevent non-specific binding. Kinases at higher concentrations that are fairly stable are often ok without BSA, but we have run into some kinases that lose activity quickly when BSA is removed, especially at lower concentrations. We have removed Brij-35 as well. Some kinases are ok without it, but others lose some activity over time.
My instrument has a monochromator, not filter cubes. Are there any changes I need to make?
When you need high sensitivity, a filter-based reader is the way to go. Monochromator-based readers are generally less sensitive because they only transmit a small fraction of light to the sample. When the monochromator selects a wavelength, most of the light from the source is lost, causing the excitation of the fluorophore to be significantly weaker. As the emitted light passes through a series of mirrors, grating and slits, light intensity is diminished even further. From the time the initial light source flashes until the specific wavelength of light finally reaches the detector, the loss of energy is compounded, which correlates with reduced sensitivity. This means you could miss a crucial result. In contrast, filter-based readers are much more efficient at delivering light to the sample, resulting in higher sensitivity. For this reason, with a monochromator, a higher enzyme concentration may be needed.
What is the upper limit of magnesium concentration?
We have titrated magnesium with a few kinases, but not all of them. The assay will tolerate up to 50 mM Mg, but the rate will likely only increase to 20 mM Mg. If increasing the Mg level, it should also be increased in the “no enzyme” blanks as well.
Have we ever run assays at different temperatures to better understand the difference in activity? Could running at RT be a significant contributor to an absence of signal?
We routinely run the assay at 30°C. Running the assay at room temperature will slow down the reaction, and running at 37°C will speed-up the reaction. The issue with running at room temperature is that this can vary widely from lab to lab. One lab might be 25°C, while another might be 20°C. This might also vary whether it is summer or winter as well. In our labs in particular, we have seen huge variations in the lab temperature. I would expect a loss in signal at room temperature, but not necessarily “no signal”.
What contaminants may be in buffers/reagents that may cause these non-linear artifacts?
Contaminants that may cause an increase in signal include metal contaminants in the reagents used to make the buffers. Adding 0.55 mM EGTA to the reaction mixture can eliminate this, especially with a 30-45 minute preincubation without enzyme.
Is DTT or TCEP more suitable for characterizing covalent? And can you replace DTT with TCEP?
Yes, you can replace DTT with TCEP! TCEP is better for characterizing covalent, Cys reactive derivatives because it does not contain thiols like DTT does. Some covalent inhibitors also contain thiols so DTT might compete for binding, while TCEP would not.
Troubleshooting Maximum Signal
Maximum RFU for the same substrate varies by kinase?
The maximum RFU for a given Sox-substrate is constant. If you see changes when using the same substrate with different kinases, this indicates that the kinase may be unstable under the conditions tested. Do a kinase titration and systematically test buffer components (e.g., add BSA or glycerol). You may also need to rule out
product inhibition.
Different Sox-based substrates have different maximum RFU?
This is expected, each individual Sox-based substrate may have a different maximum RFU as this value is influenced by the sequence of the peptide.
Plateau with constant RFU is observed
A plateau with constant RFU may indicate that sufficient time has elapsed for the enzyme to completely phosphorylate all the substrate.
Alternatively, the kinase being used is not stable under the conditions being tested. This may require you to to optimize conditions or testing kinase preparations from different suppliers.
Using a chemically synthesized phosphopeptide corresponding to the substrate provides a control to show the maximum RFU that should be achieved.
Why am I seeing less fluorescence at high ATP concentrations?
For some of the kinase titrations we do start to see inhibition at 500-1000 µM ATP. When determining ATP Km, we don’t use these points. Too high an ATP concentration either inhibits the reaction itself or results in product inhibition. Some enzymes are more sensitive to this than others.
Troubleshooting Minimum Signal
Low Signal to Background?
- The kinase concentration may be too low, the kinase is unstable, or insufficiently activated. The kinase may need to be pre-activated and/or additional co-factors may be required for the kinase to achieve full activity, e.g., Ca(II) or lipid for PKCs, Ca(II)/Calmodulin for some CAMKs. An activating kinase may also be used if it doesn't phosphorylate the Sox-based substrate.
- It may be necessary to titrate the ATP and/or Mg(II) to determine the concentration resulting in the maximum fluorescence increase, which can be peptide specific. Generally, a 1.5-fold increase in fluorescence upon phosphorylation is sufficient to achieve a robust Z' value of >0.8).
- Some samples may contain compounds that interfere with fluorescence and/or activity measurements in this assay. You may need to run a background fluorescence scan prior to kinetic data acquisition.
Below is a list of known compounds for which the indicated concentration results in < 10% inhibition of the PhosphoSens signal (higher concentrations should be avoided):
• CaCl2, 2.5 mM
• Detergents (0.01% SDS, 1% Triton X-100)
• DMSO, 10%
• DTT, 5 mM
• EDTA, 1 mM or EGTA, 2 mM
• MnCl2, 250 µM (used for Mn(II)-dependent kinases and corresponds to 250x physiological levels)
• NaCl, 150 mM
• Na3VO4, 40 µM
• β-glycerophosphate, 10 mM
Troubleshooting Background Signal
Background fluorescence increases over time?
Evaporation from wells, especially with long kinetic reads under low humidity conditions, can be significant. Without a plate seal, the reaction mixture will evaporate, and the baseline fluorescent signal will increase. Plates should be sealed with an optically clear adhesive film that still allows top-reading with minimal light scattering. We recommend DOT Scientific T480 Plate Seals.
Stay Informed
Want to hear the latest about our technology? Be among the first to learn about our latest products and services.